6.15.1 Form 6.15 Surges
R 60-1, 5.7.2.5
R 60-2, 2.10.7.7
R 60-3, 2.2.5
OIML [unit] | Test conditions surges on signal, data and control lines | Observer’s name: | |||||
Output gained | ☐ | using actual loads | |||||
Test load: | Line to line 1 kV | ||||||
☐ | simulating loading | Line to earth 2 kV | |||||
using: | |||||||
Cable: | ☐ Symmetrical line | ||||||
Date: | Start | Stop | ☐ Unsymmetrical line | ||||
Time: | Specimen: | ||||||
Ambient temperature | |||||||
Relative humidity | % | % | [unit] | ||||
Barometric Pressure | [unit] | ||||||
Cycle phase | Initial | During exposure | After | ||||
Load | |||||||
Time | Start | ||||||
Stop | |||||||
Quantity | reference | ||||||
indicated | |||||||
Error [] | |||||||
relative error [%] | |||||||
MPE [%] | |||||||
Pass | ☐ | ☐ | |||||
Fail | ☐ | ☐ | |||||
Observed faults after exposure | |||||||
Fault limit [%] | …………. | ||||||
Line to line (N/A for balanced) | Fault/Deviation | Significant | Acts on fault | ||||
↑ ▇ | ↓ ▇ | Yes | No | Yes | No | ||
3x | ☐ | ☐ | ☐ | ☐ | |||
3x | ☐ | ☐ | ☐ | ☐ | |||
Line to earth | |||||||
3x | ☐ | ☐ | ☐ | ☐ | |||
3x | ☐ | ☐ | ☐ | ☐ | |||
Observations | |||||||
Result | Pass | ☐ | Fail | ☐ | |||